9 TROUBLESHOOTING
Symptom Possible Cause Suggested Solution
The dBadge will
not switch on
The dBadge
batteries are flat
Charge the dBadge unit
(see section 4)
Incorrect time or
date is displayed
when turning on
the dBadge
The time and date
has not been set or
PC clock is set
incorrectly
Connect the dBadge to
dB35 software to set time
and date, refer to dB35
manual, ensuring PC
clock is set correctly
Calibrator is not
switched on
Switch on calibrator, see
calibrator manual
‘Cal Error’
Message
appears when
calibrating
dBadge
Calibrator is
incorrectly fitted
Push calibrator firmly over
microphone
dBadge will not
enter calibration
mode
dBadge is not
switched on or a
measurement is
taking place
Ensure dBadge is
switched on, if a
measurement run is
taking place, see section
5.4 to stop the
measurement
‘Battery Fail’
message
appears on
display
Batteries are flat Charge the dBadge, see
section 4.1
‘Memory Full’
message
appears on
display
All memory has
been used
Download measurement
runs to dB35 or clear
memory, see section 5.6
On power up and during operation, dBadge monitors hardware
and software operation and will display an error if it finds a
problem.
In the unlikely event this occurs, the dBadge will try to save any
data if it is in a measurement run. Should the error persist after
switching the dBadge off then on again, contact Casella CEL.